1. Institute of Laser Life Science, School of Biophotonics, South China Normal University, Guangzhou 510631 China;2. Key Laboratory of Laser Life Science, Ministry of Education, South China Normal University, Guangzhou 510631 China;;3. Guangzhou Key Laboratory of Spectral Analysis and Functional Probes, Guangzhou 510631 China

Fund Project:

  • 摘要
  • |
  • 图/表
  • |
  • 访问统计
  • |
  • 参考文献
  • |
  • 相似文献
  • |
  • 引证文献
  • |
  • 资源附件
  • |
  • 文章评论
    摘要:

    针对可调谐波长红外光源输出波长与实际波长存在偏差,本文设计了基于光声光谱技术的红外波长校准仪。首先,使用待校准的激光器照射装有气体小分子的光声腔,获得所装载分子的振转精细吸收光谱。然后利用这些简单气体小分子已知且精确的振转光谱峰位置与所得的光声光谱进行对比,最后,通过对比对红外激光器输出激光波长进行校准。本文利用装有甲烷和氨气气体的光声腔获得了本实验室光学参量震荡器(OPO/OPA)相应的波长校正拟合曲线,并成功对其在 2800-3600 cm-1 的范围内输出波长进行了校准。本红外光校准仪具有测量范围广,对现有的激光校准仪器有着互补的作用。

    Abstract:

    For the deviation between the actual wavelength and the output wavelength of the tunable wavelength infrared light source, an infrared wavelength calibrator based on photoacoustic spectroscopy technology was designed in this paper. Firstly, the gaseous small molecules are equipped in a photoacoustic cavity, with the tunable laser to be calibrated irradiated in the photoacoustic cavity, the vibration-rotational fine absorption spectrum of the molecule is obtained. Then, the known and accurate vibrational rotation spectrum peak positions of these simple gaseous molecules are compared with the obtained photoacoustic spectrum. Finally, the output light emitted by the infrared laser is corrected by the comparison. With this method, the existing infrared parametric oscillator (OPO/OPA,) in our laboratory has been calibrated in the range of 2800-3600 cm-1. Using a photoacoustic cavity containing gaseous methane and ammonia, the calibration fitting curve of the output wavelength of the OPO/OPA was obtained. This infrared light wavelength calibrator can be used in a wide infrared range. It is a complement to the existing laser wavelength calibrator.